Polarimetric Littrow Spectrograph - instrument calibration and first measurements

Beck, C.; Schmidt, W.; Kentischer, T.; Elmore, D.
Referencia bibliográfica

Astronomy and Astrophysics, Volume 437, Issue 3, July III 2005, pp.1159-1167

Fecha de publicación:
7
2005
Número de autores
4
Número de autores del IAC
0
Número de citas
107
Número de citas referidas
88
Descripción
We present first measurements and the calibration procedure for the Polarimetric Littrow Spectrograph (POLIS) operated at the Vacuum Tower Telescope on Tenerife, together with a brief summary of the technical characteristics of the instrument. In its present configuration, we achieve a polarimetric accuracy of about 3 × 10-3 in the visible channel (630 nm) of the instrument. The accuracy is limited by cross talk among the different polarization states. The detection limit for polarized light is about 2 × 10-3 for a 7 s exposure. Polarimetric measurements in the blue channel (Ca II H line, 396.7 nm) are strongly limited by the low photon flux. At this wavelength we present Stokes-V maps with a spatial resolution of about 0.5 arcs. The polarimetric quality of any spectropolarimeter is limited by the precision of the instrument calibration. We present a new method for self-calibration that reduces cross talk among the polarization components to 0.1%. This improvement results from a measurement of the retardance of the calibration waveplate with an accuracy of 0.1°. We demonstrate the capability of the simultaneous use of POLIS and the Tenerife Infrared Polarimeter which is integrated in the main spectrograph of the Vacuum Tower Telescope.